A laboratory fume hood is essential safety equipment that protects laboratory personnel from hazardous vapors, gases, and dusts by containing and exhausting harmful substances. Proper selection and use of fume hoods is critical for laboratory safety.
Construction: - Exhausts air to exterior via ductwork - Constant or variable air volume - Most common in research labs Advantages: - Excellent containment - Wide chemical compatibility - No filter replacement concerns Considerations: - Requires ductwork installation - Higher installation cost - Energy intensive
Construction: - Carbon filter removes contaminants - Filtered air recirculates to lab - Portable and flexible Advantages: - Easy installation - No ductwork needed - Portable - Lower operating cost Limitations: - Limited filter capacity - Not for all chemicals - Regular filter replacement required
Special Design: - Stainless steel construction - Waist-level exhaust slots - Built-in water washdown - Explosion-proof components Required For: - Perchloric acid (>70%) - Strong oxidizers - High-hazard procedures
| Application | Face Velocity (fpm) | Face Velocity (m/s) |
|---|---|---|
| General use | 80-100 | 0.4-0.5 |
| Toxic materials | 100-120 | 0.5-0.6 |
| Radioisotopes | 120-150 | 0.6-0.75 |
ASME/AIHA Standard: - Smoke visualization test - Tracer gas test (SF6) - Face velocity uniformity - Sash position testing Acceptance Criteria: - Capture velocity sufficient - No backstreaming - Adequate containment at sash opening
Ideal Location: - Away from doors and walkways - Away from air vents - Away from cross-drafts - Adequate clearance for maintenance - Accessible to services Minimum Clearances: - Sides: 6 inches (150mm) - Back: 6 inches (150mm) - Top: 12 inches (300mm)
Duct Sizing: - Calculate exhaust volume required - Size ducts for 1,500-2,000 fpm velocity - Minimize bends and length - Use smooth internal surfaces Materials: - PVC for corrosive chemicals - Stainless steel for high temperatures - FRP (Fiberglass Reinforced Plastic)
1. Keep Sash at Proper Height - Working height: 12-15 inches - Full sash only when not working 2. Keep Items Behind the Sash Plane - Minimum 6 inches from front - Do not block baffles 3. Minimize Blocked Area - Large equipment blocks airflow - Use raised stands when possible - Work in center of hood 4. Never Work with Sash Fully Open - This defeats containment purpose - Adjust sash height appropriately
Prohibited Practices: - Using fume hood as storage - Rapid arm movements in front - Placing paper in hood (airflow disruption) - Overloading with equipment - Working with incompatible chemicals
Weekly: - Clean work surface - Check for damage - Verify airflow gauge reading Monthly: - Full containment test - Clean exhaust grilles - Check ductwork connections Annual: - Professional performance testing - Calibration verification - Ductwork inspection
Laboratory fume hoods are critical safety equipment requiring proper selection, installation, and operation. Following these guidelines ensures optimal protection for laboratory personnel working with hazardous substances.